Техническая Спецификация для Rgb Lasersysteme QWAVE UV Lux-Meter, illumination measuring device, Brightness meter, DSA491006

Модели
DSA491006
Скачать
Страница из 30
Qwave User Manual 
 
16 
exposure times less than 1 second and increase slower for exposure times larger than 1 
second. If you find this to be the case, close the window by clicking on the “OK” button 
and  the  new  dark  spectrum  calibration  is  saved  into  the  device.    Otherwise  you  may 
calibrate again or click “Cancel”. 
Spectral Sensitivity 
Like in all spectrometers, the sensitivity of the Qwave spectrometer is not uniform across 
the  spectrum,  but  depends  on  the  wavelength.  This  is  called  “Spectral  Sensitivity”  and 
can be corrected by dividing each value of the spectrum by the corresponding sensitivity. 
This  sensitivity  calibration  is  also  stored  in  the  device.  However,  it  is  not  always  used 
when  taking  spectra,  because  if  the  sensitivity  is  corrected,  the  noise  level  is  unevenly 
distributed  across  the  spectrum.  Instead,  you  can  choose  to  use  the  spectral  sensitivity 
correction by selecting the “lightmeter” button in the Exposure Toolbar. 
If  you  would  like  to  recalibrate  the  spectral  sensitivity,  you  need  a  light  source  with  a 
continuous  spectrum  (like  a  tungsten  halogen  lamp)  and  a  calibrated  spectrum  for  this 
lamp. 
To  recalibrate,  take  a  spectrum  of  your  light  source  with  spectral  sensitivity  correction 
turned off. Make sure the spectrum contains as little noise as possible by using a strong 
light  source,  a  suitable  exposure  time  and  use  Temporal  Averaging.  Then  choose 
“Calibrate  Spectral  Sensitivity”  from  the  Tools  menu.  Click  on  the  “Load  reference  data” 
button  to  load  the  calibrated  reference  spectrum  (in  other  words:  a  spectrum  file  that 
shows how your light source really looks like). Based on the measured spectrum and this 
file, the software calculated the spectral sensitivity of your spectrometer and shows it as a 
dark green line. To save this new calibration to the device, click “OK”. 
If  you  also  want  to  calibrate  the  absolute  scale  of  the  intensity  axis,  you  need  to  take  a 
spectrum and also need to know the absolute optical power (in Watts) of this spectrum. If 
you measure the actual optical power with an optical powermeter, please make sure that 
the powermeter only detects wavelengths that are within the wavelength range of your 
spectrometer  (for  example  by  choosing  a  suitable  light  source).  Then  choose  “Calibrate 
power” from the Tools menu and enter the actual power. If you click “OK”, the new power 
calibration  is  stored  in  the  device  and  used  for  scaling  the  intensity  axis.  This  axis  then 
displays the spectral power density in nW/nm. 
In  addition,  you  don’t  have  to  calibrate  the  power  at  the  spectrometer  entrance.  For 
instance,  you  could  calibrate  the  power  that  enters  your  measurement  probe  or  any 
other  power  in  your  setup,  as  long  as  it  is  proportional  to  the  power  that  enters  the 
spectrometer. Just remember that the optical powers and power densities displayed by 
the software refer to this value.