MAINTENANCE 
245 
 
3.  Press the CH1 key to activate CH1. 
CH1
 
 
4.  Set the Coupling to DC from the 
bottom menu.  
 
 
5.  Set the Probe attenuation to 
Voltage, 10X
 
6.  Press the Autoset key. The 
compensation signal appears on 
the display. 
Autoset
 
 
7.  Press the Display key, then 
set the display type to 
Vector.  
Display
 
 
 
 
 
8.  Turn the adjustment point on the probe to 
make the waveform as square as possible. 
 
Under 
Compensation
Normal
Over 
Compensation