HOO12 CAN/LIN Bus Analysis
CAN Bus
LIN Bus
Bus Configuration
Bit rates
Pre-Defined or User-Select,  
100 Bit/s…4 Mb/s (HMO352x /2524),  
100 Bit/s…2 Mb/s (HMO72x…202x)
Pre-Defined or User-Select,  
100 Bit/s…4 Mb/s (HMO352x /2524),  
100 Bit/s…2 Mb/s (HMO72x…202x)
Signal Type
CAN-L or CAN-H, Single Ended or Differential Probe 
(Analog Channels only)
n/a
Sample Point Range
25…90 %
n/a
Threshold
Pre-Defined or User-Select
Pre-Defined or User-Select
Polarity
n/a
High or Low Active
Protocol Version
n/a
1.x, 2.x, J2602, 1.x or 2.x
Trigger
Source
digital Channel LCH 0…15 (Opt. HO3508),  
analog Channel CH 1…2 [CH 1…4]
digital Channel LCH 0…15 (Opt. HO3508),  
analog Channel CH 1…2 [CH 1…4]
Event
Start of Frame (SOF),  
End of Frame (EOF)
Error Frame
Error condition: 
Stuff Bit Error, CRC Error, 
Not Acknowledge, Form Error
Overload Frame 
Data Frame (11 or 29 Bit ID)
Remote Frame (11 or 29 Bit ID)
Identifier: 
0, 1, X (Don’t Care) Pattern, 
Trigger when =, ≠, <, >
Identifier and Data:
ID and 64 Bit data pattern (0, 1, X),  
trigger when =, ≠, <, >
Start of Frame (SOF),  
Wake Up Frame
Error Frame
Error condition: 
Checksum Error, Parity Error 
Synchronisation Error
Identifier: 
0, 1, X (Don’t Care) Pattern, 
Trigger when =, ≠, <, >
Identifier and Data:
ID and 64 Bit data pattern (0, 1, X),  
trigger when =, ≠, <, >
Input format
Hexadecimal or Binary
Hexadecimal or Binary
Hardware accelerated Decode
Source
digital Channel LCH 0…15 (Opt. HO3508),  
analog Channel CH 1…2 [CH 1…4]
digital Channel LCH 0…15 (Opt. HO3508),  
analog Channel CH 1…2 [CH 1…4]
Display
Bus
color coded for
Start and End of Frame:  White brackets
Data ID: 
Magenta, 
Remote ID: 
Yellow
DLC: 
White, 
Data: 
Cyan, 
CRC: 
White
ACK: 
Green, 
Overload: 
White, 
Error: 
Red
up to four lines for decoded values,
synchronous display of the Bit lines
color coded for
Start and End of Frame:  White brackets
Break: 
Magenta, 
Synchronisation: 
White
Identifier: 
Yellow, 
Parity: 
Green, 
Data: 
Cyan
Checksum: 
White, 
Error: 
Red, 
Wake Up: 
Magenta 
up to four lines for decoded values,
synchronous display of the Bit lines
Table
Display of Bus 0 or 1
Frame Number
State (Frame Type or Error Description)
Start Time,  
Identifier,  
DLC,  
CRC,  
Data
Display of Bus 0 or 1
Frame Number
State (Frame Type or Error Description)
Start Time,  
Identifier,  
Length,  
Checksum,  
Data
Format
Identifier & other: hexadecimal
Data: ASCII, binary, decimal, hexadecimal
Identifier & other: hexadecimal
Data & Checksum: ASCII, binary, decimal, hexadecimal