5.  Push  or  to view the menu. 
6.  Push / to change the value. 
7.  Push: 
   to set the change and go back to the live view. 
   or  to set the change and go back to the previous menu. 
   to cancel the change and go back to the live view. 
Note 
If the Display is set to Display All, you see the information about 
current reflected background temperature as 
BG = xx.x. 
Transmission/Transmittance Adjustment 
When you do infrared inspections through infrared-transparent windows 
(IR windows), not all of the infrared energy emitted from the objects of 
interest is transmitted through the optical material in the window. If the 
transmission percentage of the window is known, you can adjust this 
percentage in the Imager or in the SmartView  software. Adjustment of the 
transmission correction can make the accuracy of the temperature 
measurement better in many situations. 
To adjust the transmission percentage: 
1.  Push . 
2.  Push / to highlight Measurement
3.  Push  or  to view the menu. 
4.  Push / to highlight Transmission
5.  Push  or  to view the menu. 
6.  Push / to adjust the percentage.