6 Using the product
29
Material (material temperature)
Emissivity
Aluminium, bright rolled (170 °C)
0.04
Cotton (20 °C)
0.77
Concrete (25 °C)
0.93
Ice, smooth (0 °C)
0.97
Iron, emery ground (20 °C)
0.24
Iron with casting skin (100 °C)
0.80
Iron with rolling skin (20 °C)
0.77
Gypsum (20 °C)
0.90
Glass (90 °C)
0.94
Rubber, hard (23 °C)
0.94
Rubber, soft grey (23 °C)
0.89
Wood (70 °C)
0.94
Cork (20 )
0.70
Radiator, black anodised (50 °C)
0.98
Copper, slightly tarnished (20 °C)
0.04
Copper, oxidized (130°C)
0.76
Plastics: PE, PP, PVC (20 °C)
0.94
Brass, oxidized (200°C)
0.61
Paper (20 °C)
0.97
Porcelain (20 °C)
0.92
Black paint, matt (80 °C)
0.97
Steel, heat-treated surface (200 °C)
0.52
Steel, oxidized (200 )
0.79
Clay, burnt (70 °C)
0.91
Transformer paint (70 °C)
0.94
Brick, mortar, plaster (20 °C)
0.93
Reflected temperature
Using this offset factor, the reflection is calculated out due to the
low emissivity and the accuracy of the temperature measurement
with infrared instruments is improved. In most cases, the reflected
temperature is identical to the ambient air temperature. Only when
objects with strong emissions with much higher temperatures (such
Using this offset factor, the reflection is calculated out due to the
low emissivity and the accuracy of the temperature measurement
with infrared instruments is improved. In most cases, the reflected
temperature is identical to the ambient air temperature. Only when
objects with strong emissions with much higher temperatures (such