Техническая Спецификация для Extech AN310 Anemometer AN310

Модели
AN310
Скачать
Страница из 12
AN310-EU-EN V1.2   1/13 
11
Error Messages 
Error codes and solutions 
 
Air Temperature 
 
E02: Temperature reading is lower than the low range limit. Place the meter in room 
temperature for 30 minutes. If E02 continues, send the unit to Extech for repair. 
E03: Temperature exceeds the high range limit. Place the meter in room temperature for 30 
minutes. If E03 continues to appear, send the unit to Extech for repair. 
E31: The A‐to‐D converter requires replacing. Send the meter to Extech Instruments for repair. 
 
Relative Humidity (RH) 
E04: Temperature error corrupted the humidity reading. Check the temperature error solution 
section. 
E11: RH calibration error. Send to Extech Instruments for repair. 
E33: RH measurement circuit error. Send to Extech Instruments for repair. 
 
Dew Point (DP) and Wet Bulb (WBT) Temperature 
E04: Temperature or RH error. Check temperature or RH error solutions. 
 
Air velocity 
 
E03: Velocity exceeds the high range limit. Test the meter using an air velocity that is known 
to be within the specified range limits. Send the unit to Extech for repair if the E03 error 
message persists. 
 
Air volume 
 
E03: Reading exceeds the meter’s display limit. Check that the AREA setting is correct in the 
Setup Mode. 
E04: Air velocity error. Return meter to Extech for repair. 
 
Other Errors 
 
E32: Memory IC error. Reboot the meter and check it again; send it to Extech Instruments for 
repair if the error message persists. 
 
No display
 
1.  Check that the batteries are making good contact and are aligned for correct polarity. 
2.  Replace batteries. 
 
Display switches OFF automatically 
This symptom could be the normal ‘Sleep Mode’ which switches the meter OFF after 20 
minutes of in‐activity. If this is not the ‘Sleep Mode’, check whether the low battery indicator 
appears before the LCD attempts to switch ON; if yes, replace the batteries.