Calibration
2
U2300A Series DAQ Service Guide
17
Types Of Signal Sources
Ground-referenced signal sources
A ground- referenced signal source is defined as a signal 
source that is connected in some way to the building’s 
grounding system. This means that the signal source is 
connected to a common ground point with respect to the 
U2300A series DAQ (assume the host PC which is connected 
with DAQ is in the same power ground).
Floating signal sources
A floating signal source is a signal that is not connected to 
the building’s grounding system. It is also a device with an 
isolated output. Example of floating signal sources are 
optical isolator output, transformer output, and 
thermocouple.