Switching ON and OFF
To switch ON the units, press the ON/OFF button. A splash screen showing the IDEAL logo and model
identity is shown on the display. The Near-End unit attempts to detect a network and the Remote Unit.
The home screen is then shown on the display. SignalTEK II is ready for use.
identity is shown on the display. The Near-End unit attempts to detect a network and the Remote Unit.
The home screen is then shown on the display. SignalTEK II is ready for use.
To switch OFF either unit, press and hold the Power button for approximately 1/2 second, a shutdown
message is displayed on the screen. The currently stored setup is saved. If the unit does not switch OFF
within five seconds, see Master Reset. Always switch OFF the unit before removing the battery pack or
power module.
message is displayed on the screen. The currently stored setup is saved. If the unit does not switch OFF
within five seconds, see Master Reset. Always switch OFF the unit before removing the battery pack or
power module.
CAUTION
Do NOT remove the battery pack or power module when the unit is switched on.
Power Saving
Near-End Unit. Power saving preferences are selected from SETUP>SYSTEM>PREF. Auto Off can be
Disabled (unit remains ON indefinitely), or set to switch the unit OFF after three, 10 or 30 minutes of
inactivity. The backlight can be set to Always On, or to dim to 50% brightness after three minutes of
inactivity. Note that when mains power is connected the display is always on full brightness and the unit
remains ON indefinitely.
Disabled (unit remains ON indefinitely), or set to switch the unit OFF after three, 10 or 30 minutes of
inactivity. The backlight can be set to Always On, or to dim to 50% brightness after three minutes of
inactivity. Note that when mains power is connected the display is always on full brightness and the unit
remains ON indefinitely.
Remote Unit. The remote unit remains ON indefinitely whether on battery or mains power.
Master Reset
In the unlikely event of a system lock-up which prevents the unit from being switched OFF, it may be
necessary to perform a master reset. This will not delete any stored data.
necessary to perform a master reset. This will not delete any stored data.
Remove the power module or battery pack to access a small aperture in the SignalTEK II (Fig 3).
Insert a paper clip into the reset hole and press the internal reset switch.
Fig 3
Replace the power module or battery pack.