First turn-on checkout
CAUTION
Ensure that the line voltage selector switch on the 
back panel is set to match your line voltage.  Failure 
to do so could result in damage to the instrument.
Connect an appropriate IEC power cord to the DC Load and plug the power cord into an AC power 
outlet.  Ensure that nothing is connected to the INPUT terminals.  Turn the instrument on by pushing 
the Power button in.  The instrument should display 
SYSTEM SELFTEST
, then display 
0.00V   0.000A
.  A small annunciator will display 
OFF 
in the upper left corner of the display.  If you 
do not see this display, please consult the In case of trouble chapter.
The following test will verify the DC Load can operate properly in constant current mode.  Press the 
I-set button (you may have to press it twice) and use the numeric keypad to set the current to 0.1 A. 
To do this, press . (the decimal point), 1Enter.  Alternatively, you can adjust the current level with 
the knob, then click the knob or press Enter.
Connect a power supply or battery to the DC Load, observing the polarity.  Turn the power supply on 
and set a voltage of 1 volt.  On the DC Load, press the On/Off button.  You should see 
approximately 0.1 A of current flowing as shown by the instrument's ammeter and approximately 1 
volt.  The 
CC
 annunciator should be displayed, showing that the instrument is in the constant current 
mode.  
The instrument's voltage display may read a bit less than the power supply's output -- this may be 
due to a voltage drop in the wires connecting the power supply to the DC Load.  Press the   or 
▽ 
keys to see the power that is being consumed by the load.
Press the V-set button (you may have to press it twice).  Note the load is turned off.  Set the power 
supply to 10V and the supply's current to a very low level, e.g. 0.1 A.  Set the  DC Load voltage level 
well below the supply voltage level, e.g. 1 V.   Press the On/Off key.  Verify the proper voltage level 
(1 V) is shown on the display.  Press the 
△ or ▽ keys to see the power being dissipated.
Press the P-set button (you may have to press it twice).  Note the load is turned off.  Set the power 
level to 0.1 watt.   Press the On/Off key.  Press the 
△ or ▽ keys to verify that the power is about 0.1 
8500 DC Load Series                    
Version:  September 2, 2009 
Page 27 of 76