Key
Display
Enter
:TIMER STATE
Enter
Use arrow keys to set to 
:ON
.
:TIMER SET
Enter
TIMER=XXXXXS
  Enter desired time interval.  Valid values are 
1 to 60000 seconds (1000 minutes).
Enter
:TIMER SET
Esc  Esc
Standard display
Now, when you turn a load on, it will stay on for the designated time, then turn off.
To turn off timed operation, enter the menu 
:CONFIG:LOAD ON TIMER:TIMER STATE
 and set it to 
:OFF
.
Changing display resolution
The DC Load has two ranges of current and voltage (see the specifications).  You can enable higher 
resolution for voltages and currents on the lower range.  Enter the menu with Shift + Menu and set 
the menu element :CONFIG:RANGE SELECT to ON.
Increasing current resolution
To increase the current resolution (and change to the lower range), press Shift + .  If the 
sinking a current higher than the top end of the lower range, it will be reset to the top end of the 
lower range.  To change back to the lower current resolution, press Shift + 
▽ again.
Increasing voltage resolution
To increase the voltage resolution (and change to the lower range), press Shift +  .  You may see
 
OVERVOLTAGE
 message on the display if the voltage is too high.  To change back to the lower 
voltage resolution, press Shift + 
△ again.
Short
When a test mode is on, you may press Shift + Short to emulate a short.  This will draw maximum 
current from the DC supply in any of the four operation modes (CC, CV, CW, or CR).  In CC, CV, or 
CR mode, you may press  Shift + Short to stop the short.  The DC Load will return to its previous 
operation.  However, in CW mode, the short current will continue to be drawn; to stop the short, you 
must press the On/Off key
Note: When emulating a Short while in CV mode, the current software limit may trigger the DC load 
Input to turn Off, provided the current limit is reached. This does not affect the programmed settings 
and the DC load will restore the previously programmed values once the Short condition is removed.
Battery test
The battery test feature measures the time it takes for a battery voltage to drop to a specified value 
while drawing a constant current from the battery.  When the voltage at the DC Load's terminals 
reaches the specified voltage, the test is ended and the integrated current (i.e., charge supplied by 