You can download a complete python program along with detailed documentation from our website 
at 
Here are the printed results when the above script is ran:
Set to remote command:
      aa ·· 20 01 ··   ·· ·· ·· ·· ··
      ·· ·· ·· ·· ··   ·· ·· ·· ·· ··
      ·· ·· ·· ·· ··   cb
Response:
      aa ·· 12 80 ··   ·· ·· ·· ·· ··
      ·· ·· ·· ·· ··   ·· ·· ·· ·· ··
      ·· ·· ·· ·· ··   3c
The 
·
 characters represent the bytes with a value of 
0x00
.  This makes it easier to see the nonzero 
bytes in the string.
The first byte of a command is always 
0xaa
 and the second byte is the address of the DC load.  The 
address should be set to 0.  The third byte identifies the command "set to remote" and the fourth 
byte is a 1, which means enable remote mode.  If the fourth byte was 0, this command would set the 
DC load to local mode.
The third byte of the response string is 
0x12
, which means this is a packet that gives the status of 
the last command sent.  The fourth byte is 
0x80
, which means the command completed 
successfully.
On the DC load, you should see the 
Rmt
 annunciator turned on immediately after running the script. 
You will also see the 
Link
 annunciator light up while communications are going on, then blink out 
after a few seconds.
Press 
Shi f t   +  Loca l
 to set the DC load back to local mode.
We've learned two key things about the DC load:
1. Commands are always sent as 26 byte packets.
2. For any command you send to the DC load, you must also request the return of a 26 byte 
packet.  This returned packet will either be a status packet or an information packet 
containing data you requested -- for example, the power level currently set.
Get in the habit of looking at the LEDs on the IT-E131 or IT-E132 interfaces.  Every command you 
send to the DC load should result in both the RX and TX LEDs blinking once.  If this does not 
happen, something is wrong with the code, interface, or instrument.