OPERATION
81
Display
2.596
V
500.06
mA
1.2982
w
CV B Value
15.000
CV A Value
15.000
LOAD
RS232
A Value
Fine
03/Sep/2012
V
V
OCP Alarm
Please Press Enter
To Clear Alarm
OCP
OCP indicator
Alarm message
when OCP is set
to Load Off
when OCP is set
to Load Off
Mode
CV
I Range
L 0.7A
V Range
L 15V
Response
Fast
Configure
OPP
Description
For OPP, the PEL-3000 can be configured to
either limit the power or turn off the load.
either limit the power or turn off the load.
The OPP levels can be set to 10% higher than
the rating power.
the rating power.
Operation
1. Press
Main
> Configure[F5] > Protection[F1]
and set the OPP Level and OPP Setting.
Range:
OPP Level: Rating power + 10%
OPP Setting: LIMIT, Load Off
OPP Setting: LIMIT, Load Off
Alarm
When OPP Setting is configured to Load Off, a
message will be displayed on the screen when
OPP is tripped. The Enter key must be pressed
to clear the alarm message.
message will be displayed on the screen when
OPP is tripped. The Enter key must be pressed
to clear the alarm message.
When configured to LIMIT, OPP will be
displayed on the screen when the OPP is tripped
and the power will be limited to the OPP Level
setting.
displayed on the screen when the OPP is tripped
and the power will be limited to the OPP Level
setting.