Introduction 
 
The PROBE60S is a passive high impedance oscilloscope probe designed and calibrated for use 
on instruments having an input impedance of 1 Mohm shunted by 25pF. However, it may be 
compensated for use with instruments having an input capacitance of 10 - 35 pF. The probe 
incorporates a two position slide switch in the head which selects attenuation of X1, X10 position. 
 
Maintenance 
 
Before dismantling any part of the probe, make sure it is disconnected from any voltage source. 
The probe head can be detached from the cable assembly by unplugging the push fit BNC 
connector on the probe head. This permits replacement of the cable or head assembly should 
either part become damaged. The measuring tip is also replaceable. To replace a broken tip, hold 
the black insulating part of the tip with pliers and pull it away from the probe head. Replace with a 
new tip taking care to align with the inner contact. 
 
Compensation Adjustment 
 
In order to obtain the correct division ratio with each 
oscilloscope, the attenuation network needs to be 
adjusted. To compensate the probe to your 
oscilloscope, apply a 1kHz square wave to the 
probe tip, or connect to the cal socket on the 
oscilloscope to display a few cycles of the 
waveform.  
And adjust the trimmer located in the BNC connector for a flat-topped square wave. 
 
Specifications 
Position x1 
Attenuation Ratio  
1:1 
Bandwidth 
DC tot 15 MHz 
Rise 
Time 27 
ns 
Input Resistance 
1Mohm (oscilloscope input) 
Input Capacitance  
46 pF plus oscilloscope capacitance 
Working Voltage 
600 V DC incl. Peak AC derating with frequency  
Position x10 
Attenuation Ratio 
1:10 
Bandwidth 
DC tot 60 MHz 
Rise 
Time 5.5 
ns 
Input Resistance 
10 Mohm when used with oscilloscopes with 1Mohm input 
Input Capacitance  
Approx. 18pF      
Compensation Range 
10 to 35 pF 
Working voltage 
600 V DC incl. peak-AC derating with frequency (see Fig. 1)  
Cable length  
±140 cm