Measure
 
The instrument can operate in one of four electromagnetic field 
level measurement modes: 
 
● EMF in LIVE BUILDING
 
● EMF in LIVE AREA
 
● EMF of PC
 
● View 
 
"EMF in LIVE BUILDING", "EMF in LIVE AREA", and "EMF 
of PC" modes have preset thresholds for audible, color and text warning 
triggering according to the norms (Table 2).
                                                                                                           
 Table 2
                Mode 
             Electric field      Magnetic field      Averaging      
                                                threshold, V/m     threshold, uT          degree                             
EMF in LIVE BUILDING 
     500 
               10 
    10
EMF in LIVE AREA                        1000                 25 
    10
EMF of PC 
                       25 
             0,25 
    40
View  
                                         no 
              no 
    40
 
When the threshold limit is exceeded, a corresponding message is 
displayed on the red background and an interrupted alarm sounds. The signal 
can be enabled or disabled by the "Alarm" parameter setting
[Main menu] - [Settings] - [Sound] - [Alarm] 
 
The average measurement time is 500 ms. Measurement data are 
averaged (averaging degree for each measurement mode is specified in the 
Table 2). A root-mean-square value of all values along measurement axes is 
displayed as an electric and magnetic field intensity. An instant value for 
each measurement along each axis is also displayed in a form of graphic 
column (page 33). 