Справочник Пользователя для HP Water System HP VXI

Скачать
Страница из 34
30
Command Reference  
Example
*ESR?
Sends Standard Event Status Register 
query.
*GMC? <
name_string>
Get Macro Query.  Returns arbitrary block response data which contains the 
command or command sequence defined by name_string.  The command sequence 
will be prefixed with characters which indicate the number of characters that follow 
the prefix.
Example
*GMC? ‘ LIST’
Ask for definition of macro from *DMC 
example.
If the prefix returned consisted of "#214", the 2 says to expect two 
character-counting digits.  The 14 says that 14 characters of data follow.  Had the 
returned macro been shorter, such as #15*EMC?, we would read this as 1 counting 
digit indicating 5 data characters.
*IDN?
Identity.  Returns the device identity.  The response consists of the following four 
fields (fields are separated by commas):
Manufacturer
Model Number
Serial Number (returns 0 if not available)
Firmware Revision (returns 0 if not available)
The *IDN? command returns something similar to the following for the 
HP E1411B:
HEWLETT-PACKARD,E1411B,0,B,05.00
Example
*IDN?
Queries identity.
*LMC?
Learn Macros Query.  Returns a quoted string name for each currently defined 
macro.  If more than one macro is defined, the quoted strings are separated by 
commas (,).  If no macro is defined, then a quoted null string ("") is returned.
*LRN?
Learn query command.  *LRN? causes the instrument to respond with a string of 
SCPI commands which define the instrument’s current state.  Your application 
program can enter the *LRN? response data into a string variable, later to be sent 
back to the instrument to restore that configuration.
Example response from an HP E1326B voltmeter in the power-on state: