4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual
10
4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual
7/99
049
TABLE VII - OPTIONS
Tagging
None
0_
Linen
L_
Stainless Steel
S_
Certificate of Calibration
No
_ 0
Yes
_ 1
Notes:
1. 
Replacement cells only, caution look at Restrictions for insertion depth dimensions based on valve
assembly type.
2. When converting from 4806 to 4909, order 4908 directly.
3. Replacement 4908 cells for existing 4908’s and 4909’s manufactured before 8/85 must specify Table V = 02.
4. This option is application sensitive.  You must contact Analytical Instruments Marketing for approval.
RESTRICTIONS
Restriction Letter
Available Only With
Not Available With
Table
Selection
Table
Selection
c
V
Standard insertion cell for SS valve assemblies consists of a conductivity 
cell of variable length based on the cell constant and a 13.2" PES tube
molded to the cell.  The conductivity cell insertion depth is listed below:
001, X01, XX1 = 7.5"
XX5 = 6.5"
X10 = 7.7"
X20 = 8.0"
X25, X50 = 8.8"
d
V
Standard conductivity cell insertion depth dimensions according to cell  
constants for CPVC assemblies is listed below:
001, X01, XX1 = 5.5"
XX5 = 4.5"
X10 = 5.7"
X20 = 6.0"
X25, X50 = 6.8"
e
V
02
I
X25, X50
f
III
For 9782 and 7082 Analyzers only
g
I
Not for 9782 and 7082 Analyzers
h
V
02
j
V
03