Chapter 8. Servicing the PT878
180
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
8.8.1 Setting up Signal Parameters (cont.)
5.
The next prompt asks for velocity averaging.
a.
Press
[ENTER]
to open the drop-down list.
b.
Use the [
T
] or [
S
] arrow keys to scroll to the desired number of
velocity measurements to average together to smooth out noise in
the system.
the system.
c.
Press
[ENTER]
to confirm your selection.
6.
The final prompt, errors allowed, specifies the number of errors the
meter can record before displaying an error message.
meter can record before displaying an error message.
a.
Press
[ENTER]
to open the text box.
b.
Use the numeric keys to enter the desired value.
c.
Press
[ENTER]
to confirm your entry.
You have finished entering data in the Signal Parameter window. To leave
this window,
this window,
•
Press the [
T
] key to return to the Signal Parameter tab, and the [
X
]
arrow key to move to the Peak Detection tab.
•
Press
[F2]
(Cancel) to return to Operate Mode without confirming the
changes.
•
Press
[F3]
(OK) to confirm the new data. The PT878 returns to
Operate Mode.