Electromagnetic compatibility 
Warning 
When this device is connected with other electrical equipment, leakage 
currents may be additive. To minimize total leakage current per patient, 
ensure that all systems are installed according to the requirements of IEC 
60601-1-1. 
Caution 
Portable and mobile RF communications equipment may affect the normal function of 
the display. Do not use cables or accessories other than those provided with the display, 
as this may result in increased electromagnetic emissions or decreased immunity to such 
emissions. If the display is used adjacent to other equipment, observe and verify normal 
operation of the display in the configuration in which it will be used prior to using it in a 
surgical procedure. Consult the tables below for guidance in placing the display.   
 
Like other electrical medical equipment, the MD-4221 requires special precautions to 
ensure electromagnetic compatibility with other electrical medical devices. To ensure 
electromagnetic compatibility (EMC), the display must be installed and operated 
according to the EMC information provided in this manual. The display have been 
designed and tested to comply with IEC 60601-1-2:2001 requirements for EMC with other 
devices.