Руководство По Работе для Zeiss Axiovert 100
Troubleshooting
3D Microscopy
12
B 40-022-4 e 07/98
2.4
Troubleshooting
The monitor displays colored fringing or
several out-of-focus images:
several out-of-focus images:
Switch off MCU II;
switch it on again after 10 seconds
No 3D image is obtained:
monitor shutter does not function;
check switch for monitor shutter on the
control unit; the green lamp lights up only if
the switches are set correctly
control unit; the green lamp lights up only if
the switches are set correctly
parameters on the MCU II are not set
correctly;
check the parameters, mainly Time,
MEMORY IN, MEMORY OUT
check the parameters, mainly Time,
MEMORY IN, MEMORY OUT
camera not positioned correctly;
check whether the camera controls are
aligned perpendicular to the stand and
whether the monitor image follows the
stage motion;
if not, turn the camera
aligned perpendicular to the stand and
whether the monitor image follows the
stage motion;
if not, turn the camera
3D illumination shutter is out of alignment;
recenter aperture diaphragm,
recenter 3D illumination shutter
recenter 3D illumination shutter
Knob of 3D-CONTROL DAS-3 control units
not set correctly;
check 3D setting of 3D-CONTROL DAS-3
check 3D setting of 3D-CONTROL DAS-3
The image flickers:
pulse rate is not set correctly;
set 3D VIS on the 3D CONTROL DAS-3 for
observation, and 3D VIDEO for monitor
observation
observation, and 3D VIDEO for monitor
observation
A pseudostereoscopic image is displayed,
i.e. heights and depths in the image are
exchanged:
i.e. heights and depths in the image are
exchanged:
The toggle switch of the 3D CONTROL DAS-
3 is not set correctly;
set toggle switch to "STEREO"
set toggle switch to "STEREO"
For observation, exchange eyepieces for 3D
shutter
For monitor observation,
set toggle switch of 3D CONTROL DAS-3 to
"STEREO";
if required, change "Scope LCD" parameters
in the CONFIG menu of the MCU II
"STEREO";
if required, change "Scope LCD" parameters
in the CONFIG menu of the MCU II
For monitor observation,
set the switch of the control unit for the
monitor shutter to "stereo mode"
monitor shutter to "stereo mode"
Note:
If inverted microscopes are used in transmitted
light - e.g. the Axiovert 100/135/135 M -
remember that observation is made from the
opposite side of the instrument if manipulators,
needles and electrodes are used.
If inverted microscopes are used in transmitted
light - e.g. the Axiovert 100/135/135 M -
remember that observation is made from the
opposite side of the instrument if manipulators,
needles and electrodes are used.