Руководство По Работе для Zeiss Axiovert 100
3D Microscopy
Axiovert 100/135/135 M
Axiovert 100/135/135 M
Eyepiece version
B 40-022-4 e 09/98
21
2.3
Adjustment
Fig. 7
Adjustment of 3D illumination shutter,
pupil
pupil
Use cover slip as an object and focus on it
using the 10x objective;
after adjustment, check the setting using the
objective of the highest magnification
after adjustment, check the setting using the
objective of the highest magnification
2.3.1 Adjustment of 3D illumination
shutter
Move rotary switch of the 3D-CONTROL
DSA-3 to the STEP position:
Fig. 6
Fig. 6
Observe the pupil:
Axiovert 100: remove the eyepiece and
insert the centering telescope,
Axiovert 135/135 M with Bertrand lens in
the 4x turret:
use centering telescope or Bertrand lens to
focus on the aperture diaphragm;
one pupil half will be bright, the other dark:
Fig. 7
insert the centering telescope,
Axiovert 135/135 M with Bertrand lens in
the 4x turret:
use centering telescope or Bertrand lens to
focus on the aperture diaphragm;
one pupil half will be bright, the other dark:
Fig. 7
Move 3D illumination shutter to the side
until both pupil halves appear identical in
size;
if required, reset the stop screw;
Use fixation screw to hold 3D illumination
shutter in position
size;
if required, reset the stop screw;
Use fixation screw to hold 3D illumination
shutter in position
Close aperture diaphragm;
check center position