Справочник Пользователя для Siemens Sitrans F FX300

Скачать
Страница из 80
6
 OPERATION 
42 
SITRANS FX300
www.siemens.com/flow
06/2009 - SFIDK.PS.050.F2.02   A5E02100423
Example: changing the default parameter from m
3
/h to l/min
6.2.5  Changing units
When entering numbers and values in floating point format, the maximum possible accuracy is 
0.003%.  The accuracy depends on both the position of the decimal point and the length of the 
number entered.
Numerical values and factors are displayed in the first line of the 10-digit display. Numerical 
values are either displayed in floating point format (123.4567890) or are expressed in exponential 
format (123456E002). The conversion factor of the totalizer and the pulse output is, however, 
expressed in whole numbers.
Input values exponential format
Flow units
The base units are m
m
m
m
3
33
3
/h for volume flow measurement, m
m
m
m
3
33
3
/h stand.
/h stand.
/h stand.
/h stand. for standard volumetric flow 
measurement and  kg/h
kg/h
kg/h
kg/h for mass flow measurement. 
The flow units can be changed in menu item 1.1.4 "Max.Flow".
User defined units can be entered using "User Def." 
The unit (text) as well as the conversion factor (number) can be entered here.
The conversion factor must always be entered based on the base unit.
Conversion
Menu items
Procedure
Display
Procedure
Display
107.2
m
3
/h
5
^
^
^
^
0000600.00
L/min
1
3 x 
3 x 
3 x 
3 x 
1.1.1111
Language
6
^
^
^
^
Display
Unit
2
3 x 
3 x 
3 x 
3 x ↑
1.1.4444
Max. Flow
7
^
^
^
^
1111.1.4
Max. Flow
3
m
3
/h
Unit
8
^
^
^
^
Save Yes
4
3 x 
3 x 
3 x 
3 x ↑
L/min
Unit
9
^
^
^
^
1787.0000
L/min
Display position
1
2
3
4
5
6
7
8
9
10
Decimal point
-
-
-
-
-
-
Input values
0 ...
9
0 ...
9
0 ...
9
0 ...
9
0 ...
9
0 ...
9
E
-
or
0
0 ...
3
0 ...
8
To select the exponential function, the decimal point must be located between the 2nd and 5th 
decimal position.
Formula
New unit (User Def.)
(User Def.)
(User Def.)
(User Def.)
====
A1 Coeff.
A1 Coeff.
A1 Coeff.
A1 Coeff.
****
Base unit
Base unit
Base unit
Base unit
Example:
1 litre / h
=
0.001
*
m
3
/h
1.1.3
Maes.Inst
Volume / Standard volume / Mass
1.1.4
Max. Flow
Unit (User Def.) / Text / A1 Coeff.
Unit (User Def.) / Text / A1 Coeff.
Unit (User Def.) / Text / A1 Coeff.
Unit (User Def.) / Text / A1 Coeff. / Max. Flow / Flow Displ
MA_Siemens_Sitrans_FX300_R02_en_PRT.book  Page 42  Monday, June 22, 2009  8:17 AM