Operation
English
®
The chart below shows the replenishment time to be set for the requested replen-
isher rate per m
2
-film surface. The standard setting is 10 sec. with 500 ml replen-
isher rate per m
2
 film surface. The replenisher rate has to be adjusted depending
on film material, chemicals and film throughput.
Dimmer Function of Display
With the dimmer function the intensity of the control panel display can be changed.
The intensity can be reduced to avoid possible exposure of high-sensitive films.
Normal-sensitive films are not affected by the display.
Setting the intensity of the display:
1.
Switch processor on.
2.
Press “arrow down” button (3) for about 5 seconds. The intensity changes.
Two intensity-levels are available: bright and dark. During film-throughput this
function is not available.
Machine Cover Safety Disconnection
The machine cover may only be removed for service and maintenance purposes.
The processor cannot be started without the cover. In the event that the machine
cover is removed during film-throughput, the film transport will be stopped. On the
display the error message “E1” will be displayed. The error will be reset when the
machine cover has been replaced.
Replenishment time and replenishment rate relation
Replenisher rate
(ml/m
2
)
Replenisher time (s)
Replenisher rate
(ml per cycle)
200
4(4)
50
300
6(5)
75
400
8(7)
100
500
10(9)
125
600
13(11)
150
700
15(12)
175
800
17(14)
200
900
19(16)
225
1000
21(18)
250
Values in brackets for 60 Hz power connection.
Values when pump is set at 100% pump capacity.