Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Remote Operation of the Analyzer 
 
 
177 
Table 8-2: Control Input Pin Assignments 
INPUT STATUS 
CONDITION 
WHEN 
ENABLED 
A External 
Zero 
Cal 
Zero calibration mode is activated. The mode field of the display will read 
ZERO CAL R. 
B External 
Span 
Cal 
Span calibration mode is activated. The mode field of the display will read 
SPAN CAL R. 
C  
Unused 
D  
Unused 
E  
Unused 
F  
Unused 
 
Digital Ground 
Provided to ground an external device (e.g., recorder). 
DC Power For Input 
Pull Ups 
Input for +5 VDC required to activate inputs A - F. This voltage can be taken 
from an external source or from the “+” pin. 
Internal +5v Supply 
Internal source of +5V which can be used to activate inputs when connected 
to pin U. 
There are two methods to activate control inputs. The internal +5V available from the 
“+” pin is the most convenient method (Figure 8-2). However, to ensure that these 
inputs are truly isolated, a separate, external 5 VDC power supply should be used 
(Figure 8-3). 
 
CONTROL IN 
    A     B    C     D     E     F           U    
+
 
SP
A
N
ZER
O
 
 
Figure 8-2: Control Inputs with Local 5 V Power Supply 
06807C DCN6650