Remote Operation of the Analyzer 
 
 
181 
for trouble-shooting and reference purposes. Terminal emulation programs such as 
HyperTerminal can capture these messages to text files for later review. 
8.3. REMOTE ACCESS BY MODEM 
The T100 can be connected to a modem for remote access. This requires a cable 
between the analyzer’s COM port and the modem, typically a DB-9F to DB-25M cable 
(available from Teledyne API with P/N WR0000024). 
Once the cable has been connected, check to ensure that the DTE-DCE is in the correct 
position (refer to Section 6.1). Also ensure that the T100 COM port is set for a baud rate 
that is compatible with the modem, which needs to operate with an 8-bit word length 
with one stop bit. 
The first step is to turn on the MODEM ENABLE communication mode (Mode 64, 
Section  6.2.1). Once this is completed, the appropriate setup command line for your 
modem can be entered into the analyzer. The default setting for this feature is: 
AT Y0 &D0 &H0 &I0 S0=2 &B0 &N6 &M0 E0 Q1 &W0 
This string can be altered to match your modem’s initialization and can be up to 100 
characters long. 
06807C DCN6650