Calibration Procedures 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
196 
To perform a manual calibration check of an analyzer with a zero/span valve or IZS 
Option installed, use the following method: 
STEP ONE: Connect the sources of Zero Air and Span Gas as shown below.  
 
SAMPLE 
ZERO AIR 
EXHAUST 
Source of  
SAMPLE Gas 
VENT if input is pressurized 
Internal Zero/Span Option (IZS) – Option 51A 
 
Ambient 
Air 
 
 
 
 
External Zero 
Air Scrubber 
Filter 
MODEL 701  
Zero Air 
Generator 
Source of  
SAMPLE Gas 
VENT if input is pressurized 
 
MODEL 700 
Gas Dilution Calibrator 
(with O
3
  generator option) 
 
 
VENT 
 
SAMPLE 
ZERO AIR 
EXHAUST 
SPAN 1 
Calibrated 
SO
2
 gas 
(At high 
concentration)
 
VENT 
Needle valve 
to control flow
Chassis 
Chassis
 
Figure 9-9: Setup for Manual Calibration Check with Z/S Valve or IZS Option (Step 1) 
 
06807C DCN6650