Instrument Maintenance 
Teledyne API  - T100 UV Fluorescence SO2 Analyzer  
228 
11.3. MAINTENANCE PROCEDURES 
The following procedures need to be performed regularly as part of the standard 
maintenance of the T100. 
11.3.1. CHANGING THE SAMPLE PARTICULATE FILTER 
The particulate filter should be inspected often for signs of plugging or excess dirt. It 
should be replaced according to the service interval in Table 11-1 even without obvious 
signs of dirt. Filters with 1 and 5 µm pore size can clog up while retaining a clean look. 
We recommend handling the filter and the wetted surfaces of the filter housing with 
gloves and tweezers. 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
Do not touch any part of the housing, filter element, PTFE retaining ring, 
glass cover and the O-ring with bare hands, as contamination can 
negatively impact accuracy of readings.. 
To change the filter according to the service interval in Table 11-1: 
1.  Turn OFF the analyzer to prevent drawing debris into the sample line. 
2.  Open the analyzer’s hinged front panel and unscrew the knurled retaining ring of the 
filter assembly.  
 
Figure 11-1: 
Sample Particulate Filter Assembly 
06807C DCN6650