Troubleshooting & Service 
Teledyne API  - T100 UV Fluorescence SO2 Analyzer  
250 
 
J2: Power 
Configuration 
Jumper 
 
Figure 12-4: 
Location of Relay Board Power Configuration Jumper 
AC Configuration of the pump is accomplished via an in-line, hard wired, set of 
connections.  Call Teledyne API’s Technical Support Department for more information.  
12.6.2. DC POWER SUPPLY 
If you have determined that the analyzer’s AC main power is working, but the unit is 
still not operating properly, there may be a problem with one of the instrument’s 
switching power supplies, which convert AC power to 5 and ±15 V (PS1) as well as +12 
V DC power (PS2). The supplies can either have DC output at all or a noisy output 
(fluctuating). 
To assist tracing DC Power Supply problems, the wiring used to connect the various 
printed circuit assemblies and DC powered components and the associated test points on 
the relay board follow a standard color-coding scheme as defined in Table 12-4. 
Table 12-4: DC Power Test Point and Wiring Color Code 
NAME TEST 
POINT#  COLOR 
DEFINITION 
DGND 1  Black 
Digital 
ground 
+5V 2  Red 
 
AGND 3  Green  Analog 
ground 
+15V 4  Blue 
 
-15V 5 Yellow 
 
+12V 6  Purple 
 
+12R 
Orange 
12 V return (ground) line 
A voltmeter should be used to verify that the DC voltages are correct as listed in Table 
12-4. An 
oscilloscope, in AC mode and with band limiting turned on, can be used to 
evaluate if the supplies are excessively noisy (>100 mV peak-to-peak). 