Teledyne API - T100 UV Fluorescence SO2 Analyzer  
SETUP Menu 
 
 
95 
5.4.3.3. AUTO RANGE MODE (AUTO) 
In AUTO range mode, the analyzer automatically switches the reporting range between 
two user-defined ranges (low and high). The unit will switch from low range to high 
range when either the SO
2
 concentration exceeds 98% of the low range span. The unit 
will return from high range back to low range once both the SO
2
 concentration falls 
below 75% of the low range span. 
In AUTO Range mode the instrument reports the same data in the same range on both 
the  A1 and A2 outputs and automatically switches both outputs between ranges as 
described above. Also, the RANGE test function displayed on the front panel will be 
replaced by two separate functions: 
  RANGE1: The LOW range setting for all analog outputs. 
  RANGE2: The HIGH range setting for all analog outputs. 
The high/low range status is also reported through the external, digital status bits (refer 
to Section 8.1.1).  
To set individual ranges press the following control button sequence. 
 
 
 
 
 
 
SAMPLE          RANGE = 500.000 PPB        SO2 =XXX.X     
 
 < TST  TST >  CAL  
 
 
SETUP
 
 
 
 
SETUP X.X 
 
      
PRIMARY SETUP MENU
 
  
 
 CFG  DAS
  RNGE  
PASS  CLK  MORE  
EXIT
 
 
 
 
SETUP X.X 
RANGE CONTROL MENU 
 
 
 
 
MODE
 
 
 
 
 
 
SET  UNIT
 
 
 
 
EXIT
 
 
 
 
SETUP X.X 
RANGE MODE: 
SNGL
 
 
 
 
 SNGL  IND   
 AUTO
 
 
     ENTR  EXIT
 
SAMPLE          
ENTER SETUP PASS : 818
 
 
        
 8        1        8 
 
                      
ENTR
  
 EXIT
 
EXIT
 x 2 returns 
to the main 
SAMPLE
 display  
 
 
 
SETUP X.X 
RANGE MODE: 
AUTO
 
 
 
 SNGL
 
 
 
 
IND    AUTO 
 
     ENTR
  EXIT
 
 
 
 
SETUP X.X 
RANGE CONTROL MENU 
 
 
 MODE
 
 
 
 
SET
  UNIT
 
 
 
 
EXIT
 
 
 
 
SETUP X.X 
HIGH RANGE: 
500.0 Conc
 
 
 
 0          0         5          0          0          .0
       
ENTR
 
 EXIT
 
 
 
 
SETUP X.X 
LOW RANGE: 
500.0 Conc
 
 
 
 
0          0         5          0          0          .0 
     
ENTR
 
 EXIT 
Toggle the numeral 
buttons to set the 
LOW
 and HIGH 
range value. 
ENTR accepts the 
new setting, EXIT 
ignores the new 
setting. 
 
Figure 5-6: SETUP RNGE – Auto Range Mode 
 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
In AUTO range mode, the LOW and HIGH ranges have separate slopes 
and offsets for computing SO
2
 concentration. The two ranges must be 
independently calibrated. 