Справочник Пользователя для National Instruments Welding System 321645c-01

Скачать
Страница из 554
Chapter 2
Function Reference — LPM16_Calibrate
©
 National Instruments Corporation
2-279
NI-DAQ FRM for PC Compatibles
LPM16_Calibrate
Format
status = LPM16_Calibrate (deviceNumber)
Purpose
Calibrates the LPM devices converter. The calibration calculates the correct offset voltage for 
the voltage comparator, adjusts positive linearity and full-scale errors to less than ±0.5 LSB 
each, and adjusts zero error to less than ±1 LSB.
Parameters
Input
Using This Function
When the function is called, the ADC1241 ADC goes into a self-calibration cycle. The 
function does not return until the self-calibration is completed.
Name
Type 
Description
deviceNumber
I16
assigned by configuration utility