Справочник Пользователя для Honeywell Work Light DR4300

Скачать
Страница из 262
DR4300 Circular Chart Recorder 
80 
DR4300 Circular Chart Recorder Product Manual 
12/03 
4.3.5.2  Setting SW6 Input Switches 
SW6 switches 1, 3, 4, 5, 6 define input type 
In a recorder model having a display and keypad the input switches at SW6 must be set to match the type 
of input to be used. The location of SW6 is shown in Figure 4-3. The appropriate switch settings for the 
various input types are shown in Table 4-3.  
Note that the setting of switch 2 in SW6 is not related to input type. It is used, along with the “BRNOUT” 
parameter in the input group, to specify the effects of an input failure as described in Subsection 4.3.5.3. 
 
ATTENTION 
The ON position for SW6 switches is toward the left side of the printed circuit assembly (when 
facing recorder.) 
 
Table 4-3  SW6 Input Switch Settings for Models  
Having Display and Keypad 
Input Type 
SW6 Input Switch Settings 
 
1 2 3 4 5 6 
0 V to 1 V 
off 

 
off 
ON 
off off 
0 V to 2 V 
off 

 
off 
ON 
off off 
0 V to 5 V 
off 

 
off 
ON 
off off 
0 V to 10 V 
off 

 
off 
ON 
off off 
2 V to 10 V 
off 

 
off 
ON 
off off 
1 V to 5 V 
off 

 
off 
ON 
off off 
0 mV to 10 mV 
off 

 
off off ON 
off 
0 mV to 100 mV 
off 

 
off off ON 
off 
0 to 200 mV 
ON 

 
off off off off 
0 mA to 20 mA 
off 

 
off 
ON 
off 
ON 
4 mA to 20 mA 
off 

 
off 
ON 
off 
ON 
all thermocouples 
off 

 
off off ON 
off 
all RTDs 
off 
off 
ON 
off 
ON 
off 
 
SW6 Input Switch 
6
5
4
3
2
1
ON
 

 This switch is used to select whether the input value used by the 
recorder should be driven over range or under range in case of input 