  For Safe Handling 
SAFETY - 12
 
 
Precautions for Proper Operation 
 
―  Precautions for the Proper Operation of the S-3400N  ― 
 
About Electricity 
 
(a)  Single-phase AC 100 V(
±10%), 50/60 Hz, 2 kVA (for main 
unit).When the supplied power line voltage is 100 V, the 
instrument works by direct connection. 
When the supplied power line voltage is other than 100 V (200 V 
or others), use a step down transformer to step the power line 
voltage down to 100 V. Refer to Chapter 1 (Specifications and 
Installation Requirement)
 for the step down transformer 
requirement. 
(b)  Connect the instrument to ground based on the standard of your 
country to prevent electric shock hazards due to the high 
voltage. Otherwise, you could suffer an electric shock or a 
malfunction could occur in the instrument. 
Avoid sharing the ground terminal with other power equipments. 
Be sure to ground the instrument independently. The image 
trouble and the malfunction of the equipment might be caused 
when sharing. 
 
Handling Volatile Solvents 
 
When handling a volatile solvent in cleaning components of this 
system, take precautions on the following points: 
(a)  Volatile solvent should be handled at a place which is well 
ventilated and is well away from flames. 
(b) For handling volatile solvents, ventilation equipment and 
protective gear should be provided at the customer’s discretion.   
The inhaling of a volatile solvent in excessive amounts can 
results in respiratory difficulties.    If volatile solvent is swallowed 
or it gets onto the skin or into the eyes, symptoms of polyneuritis 
including anesthesia and ataxia (difficulty in walking) may 
develop.