3.8  Shutting Down 
3 - 114 
(3)  If the vacuum mode is set to Low Vacuum (VP-SEM) on the Operation Panel, change it to 
High Vacuum (SEM). 
 
(4)  Grasp the handle. Insert the specimen stage into the specimen chamber.    On the EVAC 
panel on the front side of the main unit, press the EVAC button.  Alternatively, click EVAC 
located in the upper right section of the Control Panel.    The evacuation of the specimen 
chamber commences. 
If the specimen chamber had already been in a fairly good vacuum condition, a high vacuum 
state can be attained in approximately 2 minutes. 
 
 
 
 
 
 
EVAC panel 
Control Panel 
 
Fig. 3.8-5    Evacuating the Specimen Chamber 
 
  Caution:  When inserting the specimen stage into the specimen chamber, be careful that 
your fingers are not caught between the stage and the specimen chamber. 
 
 
3.8.4    Closing Windows and Shutting off the Display Power 
 
Close Windows.    Turn off the power. 
 
(1)  On the S-3400N Main Window, either click the Close button 
 or select Exit from the File 
menu.    A termination confirmation message will appear: 
 
 
 
Fig. 3.8-6    Termination Confirmation Message