Страница из 466
3.9.6 
3 - 129 
(2)  Loading a Set of Operating Conditions 
Use the following steps to load a set of operating conditions. 
 
(a) Open the Setup dialog window   
Select Condition Load from the Operate menu, or   
click the Setup button in the tool button section of the 
Operation panel and select Op.Cond tab.   
Select Load Memorized Condition on it. 
 
(b)  Select a condition name 
Memorized condition names are listed. Select a name 
by clicking it. 
Use sample picture and comment for reference of 
selection. 
 
(c)  Select items to be loaded 
Check items ( Vacc to Contrast) you wish to load and 
reproduce.  Stage position is effective only for Type 2 
instrument. 
 
(d)  Reproduce column condition 
Click the Load button.    Memorized data are retrieved 
and the electron optical column is set.   
 
If the Stigma/Alignment is included, a message will 
appear for confirmation.    These are possibly changed if 
the loaded condition was saved long before.     
In such cases, it may be better not to reproduce Stigma and Alignment data.    The 
recent data is memorized automatically y abnother condition memory function   
(the most recent condition for each of Vacc - Condenser lens setting is memorized and 
reproduced automatically).   
 
(3)  Deleting an operating condition   
Select a condition name in the same manner as above, and click the Delete button.   
 
 
3.9.7  Pseudo Color Display 
 
Use function of SEM Data Manager for pseudo color imaging. 
See 3.11.3.7  Contrast Conversion. 
 
 
 
Fig. 3.9-21  Op.Cond  
Window (call)