Troubleshooting
Proline Promass 80 PROFIBUS PA
76
692
S: SIM. MEASURAND
!: # 692
0x60
0x61
0x62
UNCERTAIN
Simulated 
Value
(manually 
specified 
value)
O.K.
Low
High
Simulation 
measured value 
active
Cause:
Simulation is active
Remedy:
Switch off simulation:
Access:
SUPERVISION 
→ SYSTEM → OPERATION → 
SIM. MEASURAND (
→ OFF)
698
S: DEV. TEST ACT.
!: # 698
0x60
0x61
0x62
UNCERTAIN
Simulated 
Value
(manually 
specified 
value)
O.K.
Low
High
Device test via 
Fieldcheck active
Cause:
The measuring device is being checked on site via 
the test and simulation device.
No.
Device status message
(local display)
PROFIBUS measured value status
Extended 
diagnostic 
message in the 
PROFIBUS 
master
Cause/remedy (spare part 
→ Page 80 ff.)
Q
u
a
lit
y co
de (H
EX
)
M
e
asured va
lue st
atus
Qua
li
ty st
a
tus
Qu
ality
 substatus
Li
m
its