Instruction 0011-9026 
Page 5
Figure 5
Figure 5a
Figure 5b
Figure 6
CO
2
  FYRITE  OPERATION  
 
 
   
 
O
2
  FYRITE  OPERATION
        (RED FLUID)   
 
 
 
 
   
 
 
       (BLUE FLUID)
5.  Holding FYRITE upright 
(Fig. 5), loosen locknut at 
rear of scale. Slide scale 
(Fig. 5a) until top of fl uid 
column lines up with zero 
line on scale (Fig. 5b). 
Tighten scale locknut.
  NOTE: When setting 
scale zero, hold FYRITE 
vertically as shown and 
level with eyes while 
sighting across scale to 
top of fl uid column.
6.  To pump gas sample 
into FYRITE (Fig. 6), 
insert open end of metal 
sampling tube into area 
bearing gas for analysis. 
Hold FYRITE in upright 
position and place sam-
pling assembly rubber 
connector tip over the 
Plunger Valve.
  Depress Plunger Valve 
fi rmly with connector tip.
  Pump sample by squeez-
ing and releasing as-
pirator bulb 18 times. 
During 18th bulb squeeze 
(with bulb held defl ated) 
release connector tip and 
Plunger Valve.
5.  Holding FYRITE upright 
(Fig. 5), loosen locknut at 
rear of scale. Slide scale 
(Fig. 5a) until top of fl uid 
column lines up with zero 
line on scale (Fig. 5b). 
Tighten scale locknut.
  NOTE: When setting 
scale zero, hold FYRITE 
vertically as shown and 
level with eyes while 
sighting across scale to 
top of fl uid column.
6.  To pump gas sample 
into FYRITE (Fig. 6), 
insert open end of metal 
sampling tube into area 
bearing gas for analysis. 
Hold FYRITE in upright 
position and place sam-
pling assembly rubber 
connector tip over the 
Plunger Valve.
  Depress Plunger Valve 
fi rmly with connector tip.
  Pump sample by squeez-
ing and releasing as-
pirator bulb 18 times. 
During 18th bulb squeeze 
(with bulb held defl ated) 
release connector tip and 
Plunger Valve.