Справочник Пользователя для Barco DICOM THEATER MGP 15

Скачать
Страница из 102
Optical sensor connection
80
Optical sensor connection
The optical sensor can be attached to the screen by means of 
Velcro, attached to the rear of the sensor. 
The sensor must be connected to the PC through USB.
Note: When you plug in the optical sensor to the PC USB bus for 
the first time, you will be asked to install a driver. You can find 
the driver on the Dicom Theater application CD-ROM, in the folder 
Drivers > I-Guard.
To connect the sensor:
1. Before attaching the sensor, be sure to set the sensor patch 
to the desired location (see above). You can always change 
the position of the patch by means of MediCal Pro.
2. Peel off the protective sheet of the Velcro part that will be 
attached to the screen. 
Figure 28
3. Attach the sensor to the screen, in the center of the optical 
sensor patch.
4. Connect the end of the sensor cable to a free USB connector 
on the PC.
Should the USB cable be too short, you will have to use the 
USB extender cables supplied with the system (2 cables per 
sensor, each 5m long).
5. If the system asks you to, install the driver for the optical 
sensor (Sequel driver).