Using TheaterWatch
MGP D5 Dicom Theater Lite
87
To adjust display luminance manually, adjust the Luminance 
slider. This function is not available for MFCD/MFGD 1218 
displays and MGP D5 projectors.
The luminance setting is saved when the system is shut 
down.
If the slider is positioned above the CAL mark, the luminance 
is in calibrated position, as determined during the latest 
calibration. If the slider is not above the CAL mark, the 
luminance is not in calibrated position.
Important
If you want the system to be DICOM compliant, you must select 
the DICOM display function and calibrate the display. After 
calibration, the luminance must remain in calibrated position to 
maintain DICOM compliance.
MGP D5 projectors are always automatically calibrated to the 
DICOM display function.