Справочник Пользователя для Barco Coronis Fusion 10MP

Скачать
Страница из 80
34
Coronis Fusion 10MP
Reading rooms
A
BOUT
 
READING
 
ROOMS
The American Association of Physicists in Medicine (AAPM) composed a 
list of pre-defined reading rooms. Each of these reading rooms are 
defined by following parameters:
the maximum light allowed in this type of room
the preset ambient light correction value for this reading room
These parameters are stored in your display and determine the preset 
ambient light correction value to take into account to recalculate the 
DICOM display function when Ambient Light Compensation (ALC) is 
enabled. Please refer to “Ambient Light Compensation (ALC)” on 
page 33 t
o enable ALC.
The available reading rooms for your C
ORONIS
 F
USION
 10MP display are:
Reading rooms can only be selected when the DICOM display 
function is selected. Therefore, please refer to “Display 
functions” on page 32 
to correctly set the display function.
CR/DR/MAMMO
Corresponds to light conditions in 
diagnostic reading rooms for computed 
radiology, digital radiology or 
mammography. This setting has the lowest 
maximum ambient light.
CT/MR/NM
Corresponds to light conditions in 
diagnostic reading rooms for computed 
tomography, magnetic resonance or 
nuclear medicine scans.
Staff Office
Corresponds to light conditions in office 
rooms.
Clinical Viewing 
Room
Corresponds to light conditions in 
diagnostic reading rooms for clinical 
viewing.