DICOM .....................Select a DICOM lookup table for most medical 
viewing applications. The DICOM function 
results in more visible grayscales in the images. 
CRT ..........................Select a CRT (gamma) lookup table in case the 
display is used to replace a CRT display. 
Apart from these factory lookup tables, there is a User programmable 
lookup table. It is default filled out with the DICOM lookup table 
(dedicated software needed to re-program it). 
Filter* 
* Not in DVI mode 
 
The display electronics provide 4 different low-pass filters to filter 
possible high-frequent distortion from an analog video signal. 
Select a different filter in case the image contains high-frequent noise 
or distortion. Select the filter that gives the best result. 
Re-calibration of the phase is necessary after selection of a different 
filter. 
Power Led 
This setting allows you to switch off the power LED.  
A check mark (3) means it is switched on, a cross (2) means it is off. 
Note: 
The LED’s orange DPMS state is not influenced by this setting. So, 
when the display goes into power-saving mode, the LED will turn 
orange, even if it was switched off by this setting. 
ALC active 
This setting allows you to switch ALC (Ambient Light Compensation) 
off or on. 
The ALC system automatically adapts the display light output, 
depending on the ambient light in the room. The ambient light is 
measured by the ALC optical sensor located at the front of the 
display. 
DPMS active 
This setting allows you to switch DPMS (Display Power Management 
System) off or on 
Orientation**  
** Only when Portrait Accelerator is installed 
Select the desired option: