Справочник Пользователя для Philips HEARTSTART XLT M3500B
Specifications
13-2
Specifications & Safety
Specifications
Defibrillator
Waveform: Biphasic Truncated Exponential. Waveform parameters adjusted
as a function of patient impedance.
as a function of patient impedance.
Shock Delivery: Via multifunction defib electrode pads.
Charge Time: Less than 3 seconds to 200 Joules with a new, fully charged
M3516A SLA battery pack at 25
o
C. Less than 15 seconds to 200 Joules when
powered by a M3517A or M3518A power module with no battery installed.
Patient Impedance Range: 25 to 180 Ohm.
Manual Mode
Manual Output Energy (Delivered): 5, 10, 25, 50, 70, 100, 150, 200
Joules.
Joules.
Controls: On/Energy Select, Off, Charge/Disarm, Shock, ECG Lead
Select, SpO
Select, SpO
2
, HR Alarms, Manual Mode, Sync, Pacer, Pacer Start/Stop,
Pacer Rate, Pacer Current, Pacer Mode, Display Contrast, ECG Gain, Vol-
ume, Print Strip, Print Summary, Annotated Mark Event.
ume, Print Strip, Print Summary, Annotated Mark Event.
Indicators: LCD display for ECG waveform and text prompts, Audio
alerts, QRS Beeper, Charging Tone, Charge Done Tone, Manual Mode
LED, Sync LED, Pacer LED, Printer.
alerts, QRS Beeper, Charging Tone, Charge Done Tone, Manual Mode
LED, Sync LED, Pacer LED, Printer.
Armed Indicators: Charge done tone and available energy indicated on
display.
display.
Energy Selection: Front Panel "1" key.
Charge Control: Front Panel "2" key.
Shock Control: Front Panel "3" key.
Synchronizer: SYNC message appears on the monitor and is annotated on
the printer (if printing while in Sync Mode). An audible beep sounds with
each detected R-wave, while a tick mark on the monitor and printed strip
indicate the discharge points. Synchronizer delay is less than 60 msec from
peak R-wave to peak current of the defibrillation discharge.
the printer (if printing while in Sync Mode). An audible beep sounds with
each detected R-wave, while a tick mark on the monitor and printed strip
indicate the discharge points. Synchronizer delay is less than 60 msec from
peak R-wave to peak current of the defibrillation discharge.