Defibrillating
6-5
6
Figure 6-4  Charging Display
If desired, you may increase or decrease the selected energy level after 
pressing the 
 button. To do so, press 
 repeatedly until the 
desired energy level is displayed. The defibrillator charges to the selected 
energy automatically. Wait until the current charge reaches the selected 
energy level before proceeding. 
3. Shock
Confirm that a shock is still indicated. Make sure no one is touching the 
patient or anything connected to the patient. Call out “Clear.” Then press 
 to deliver a shock to the patient.
To disarm the defibrillator, press 
. If 
 is not pressed in 30 sec-
onds, the defibrillator disarms automatically. 
If additional shocks are indicated, repeat the defibrillation process
WARNING
Defibrillation current can cause operator or bystander injury. Do not touch the 
patient, or equipment connected to the patient, during defibrillation.
ENERGY
SELECT
HR ALARM
LEAD
SELECT
DISARM
Shocks:0
00:00:8
78
Pads
SpO2
150 J
87 J
Current Charge
&+$5*(
(1(5*<
6(/(&7
6+2&.
',6$50
6+2&.