Thermo Scientific Revco Cryogenic Storage Freezers
Installation and Operation
4
6.2 Start Up
Refer to Section 6.1 and Figure 2 on page 3 as you complete the 
following procedures.
6.2.1 Turning the Power On
To start up your cryogenic freezer, complete the following steps:
1. Plug the freezer into the power outlet (refer to Section 4.2 on 
page 2).
2. Turn the key switch to the POWER ON position. The Power 
On LED and the Cabinet Temperature LED illuminate.
Note:
The alarm function is not active at this time. Refer to 
Section 5.2.3 for information about setting the alarms. In 
addition, the Setpoint Security will be deactivated in this 
key position.
3. The unit will start about 30 seconds later.
6.2.2 Setting the Cabinet Temperature
To set the cabinet temperature, complete the following steps:
1. Press and hold the Control Setpoint pad.
2. The Control setpoint LED lights and the Cabinet 
Temperature LED goes out.
3. Press and hold 
 to increase the temperature or 
 to 
decrease the temperature. The digital display scrolls through 
temperature settings.
4. Release both pads when the digital temperature display 
window shows the correct setpoint value.
Note:
If no keys are pressed within ten seconds, the 
temperature display reverts to the cabinet temperature.
6.2.3 Setting the Alarms
To set the cold alarm, complete the following steps:
1. Confirm that the key switch is at the Power On position 
(Setpoint Security disabled) 
2. Press and hold the Cold Alarm Setpoint pad. The LED next 
to this pad lights. The temperature display shows the Cold 
Alarm value.
3. Press and hold 
 or 
 to adjust the Cold Alarm Setpoint.
4. Release both pads when the digital temperature display 
window shows the correct setpoint value.
To set the warm alarm, complete the following steps:
1. Confirm that the key switch is at the Power On position 
(Setpoint Security disabled).
2. Press and hold the Warm Alarm Setpoint pad. The LED next 
to this pad lights. The temperature display shows the Warm 
Alarm value.
3. Press and hold 
 or 
 to adjust the Warm Alarm Setpoint.
4. Release both pads when the digital temperature display 
window shows the correct setpoint value. 
When the cabinet temperature drops below the Warm Alarm 
setting, turn the key switch to ALARM ON. The freezer is ready 
to operate.
Note:
If a power failure lasting over 30 seconds occurs, the 
POWER ON light goes out and the POWER FAILURE 
light and digital temperature display flash 
simultaneously.
WARNING!
You must turn the three-position key 
switch to the ALARM ON/SETPOINT SECURITY 
position to activate the alarm and place it in security 
operation.
6.2.4 Alarm Reset & Status Lights 
The alarm reset feature ensures user acknowledgment of the 
occurrence of certain alarm conditions. This provides greater 
monitoring power by alerting the user if an alarm condition has 
occurred during periods when the freezer must be left 
unattended. The alarm conditions incorporating this feature are:
Temperature Failure-Warm Alarm
Temperature Failure-Cold Alarm
Voltage Low
Power Failure
Life-Guard Alert
Extreme Ambient Alert 
During any of these alarm conditions, the corresponding 
indicator will flash quickly, approximately 90 times per minute. 
If the alarm condition then disappears, the flash rate will 
decrease to approximately 15 times per minute. The indicator 
will then remain flashing at this rate, unless the condition 
reoccurs, until the Alarm Reset button is pressed. Once this has 
occurred, the indicator will no longer be lit.
In the case of TEMPERATURE FAILURE-WARM ALARM and 
TEMPERATURE FAILURE-COLD ALARM, if these 
conditions have occurred but are now within the given limits, the 
two indicators will alternately flash (180° out of phase). In 
addition, for these two temperatures failures, the highest 
temperature and lowest temperature, respectively, during the 
error condition will be saved. Prior to the temperature failure 
being reset, the temperature extreme may be viewed by pressing 
the Cabinet Temperature pad in conjunction with either the 
Warm Alarm Setpoint pad or the Cold Alarm Setpoint pad.
Note:
Once the Alarm Reset button is depressed, the extreme 
temperature value is reset to the current cabinet 
temperature.
6.2.5 Alarm Test 
To test the cold and warm alarms, complete the following steps:
1. Press and hold the Alarm Test pad. The digital temperature 
display indicates rising temperature. When the temperature 
reaches the warm alarm value, the alarm sounds.
Note:
The temperature of the refrigerated space does not 
change during this procedure. Only the sensor is heated.
2. Release the Alarm Test pad, the sensor and display return to 
cabinet temperature in a few minutes.