Chapter 4 
CDMA Measurements - Screens and Control Fields
Gen Special
Input Atten
Gen Special
Generator special modes change how IQ modulated signals are 
generated.
Normal
 sets the CDMA generator and CDMA analyzer to create and 
analyze signals with standard rotation. 
Inverted
 sets reverse rotation for I and Q signals; opposite to the 
direction specified in the IS-95 standards. In this mode, the CDMA 
generator creates reverse-rotation IQ signals and the CDMA 
analyzer expects reverse-rotation IQ signals.
Screens on which this field is present
CDMA GENERATOR
Input Atten
This field controls the RF attenuators of the Test Set. If an 
ADC 
Overdriven
 message is displayed at the top of the screen while 
making measurements, increase the attenuator setting until the 
message is no longer displayed. This is the coarse level adjustment for 
the signal level into the CDMA analyzer. Fine level adjustment is 
performed using the 
Gain
 field. 
See the 
 for more information.
NOTE
The 
Input Atten
 setting does not affect, and is not 
displayed with, the average power measurement.
Screens on which this field is present
CDMA ANALYZER
CODE DOM (Code Domain Analyzer): gain controls menu
Main Menu