Справочник Пользователя для HP Cell Phone Accessories E6380-90019
102
Chapter 4
CDMA Measurements - Screens and Control Fields
Gen Special
Input Atten
Gen Special
Generator special modes change how IQ modulated signals are
generated.
generated.
•
Normal
sets the CDMA generator and CDMA analyzer to create and
analyze signals with standard rotation.
•
Inverted
sets reverse rotation for I and Q signals; opposite to the
direction specified in the IS-95 standards. In this mode, the CDMA
generator creates reverse-rotation IQ signals and the CDMA
analyzer expects reverse-rotation IQ signals.
generator creates reverse-rotation IQ signals and the CDMA
analyzer expects reverse-rotation IQ signals.
Screens on which this field is present
CDMA GENERATOR
Input Atten
This field controls the RF attenuators of the Test Set. If an
ADC
Overdriven
message is displayed at the top of the screen while
making measurements, increase the attenuator setting until the
message is no longer displayed. This is the coarse level adjustment for
the signal level into the CDMA analyzer. Fine level adjustment is
performed using the
message is no longer displayed. This is the coarse level adjustment for
the signal level into the CDMA analyzer. Fine level adjustment is
performed using the
Gain
field.
See the
for more information.
NOTE
The
Input Atten
setting does not affect, and is not
displayed with, the average power measurement.
Screens on which this field is present
CDMA ANALYZER
CODE DOM (Code Domain Analyzer): gain controls menu
CODE DOM (Code Domain Analyzer): gain controls menu
Main Menu