Chapter 5 
Analog Measurements - Screens and Control Fields
Detector
Detector
Detector 
This setting selects the type of detector used when measuring and 
displaying AF signal levels. 
Detector Types
RMS
 displays the root mean square (rms) value of signals. 
RMS*SQRT2 
displays the rms value of a signal multiplied by 
Pk+
 displays the positive peak value. 
Pk
 displays the negative peak value. 
Pk
±
/2
 adds the positive and negative peak values, and divides the 
sum by 2. 
Pk
±
Max
 compares the positive and negative peaks and displays the 
greater value (polarity is not indicated). 
Pk+ Hold
 displays and holds the positive peak value until the 
measurement is reset. To reset, use 
Meas Reset
 (press and release the 
Shift
 key, then the 
Hold
 key), select a different detector, or re-select the 
same detector. 
Pk
 Hold
 displays and holds the negative peak value until the 
measurement is reset. To reset, use 
Meas Reset
 (press and release the 
Shift
 key, then the 
Hold
 key), select a different detector, or re-select the 
same detector. 
Pk
±
/2 Hold
 divides the sum of the positive and negative peak 
values by 2, and displays the value until the measurement is reset. 
To reset, use 
Meas Reset
 (press and release the 
Shift
 key, then the 
Hold
 
key), select a different detector, or re-select the same detector. 
Pk
±
Mx Hold
 compares the positive and negative peaks and displays 
the greater value until the measurement is reset. To reset, use 
Meas 
Reset
 (press and release the 
Shift
 key, then the 
Hold
 key), select a 
different detector, or re-select the same detector. 
HP-IB Example 
Operating Considerations
 
Screens on which this field is displayed
AF ANALYZER
2