Chapter 5 
Analog Measurements - Screens and Control Fields
Port/Sweep
Port/Sweep
Port/Sweep
This control performs two functions: 
• The upper field specifies the output port of the tracking generator.
• The lower field specifies whether the tracking generator sweeps from 
low-to-high frequencies (
Norm
), or from high-to-low frequencies 
(
Invert
). (The spectrum analyzer always sweeps from low to high 
frequencies.) The swept frequency range is determined by the 
Span
 
setting in the spectrum analyzer’s main controls menu.
This field is displayed only when 
Track
 is selected in the 
Track/Fixed
 
field on the spectrum analyzer’s 
RF Gen
 controls menu.
Operating Considerations
When using the tracking generator, if the output port is set to 
RF Out
or the main menu 
Input Port
 is set to 
RF In
, internal instrument 
coupling can occur. For the best isolation between the tracking 
generator and the spectrum analyzer, use 
Dupl
 for the output, and 
Ant
 
for the input. 
For measurements on high-power devices, such as amplifiers, use the 
RF IN/OUT port for the input.
Screens on which this field is displayed
SPEC ANL: RF generator controls menu
Main Menu