Configuration - Screens and Control Fields
Abort Print
Antenna In
Control Fields on the Configuration Screens
Abort Print
Select this field to interrupt the printing job in progress. 
Abort Print is not directly programmable over the HP-IB.
Screen(s) Where Field is Present
PRINTER CONFIGURE
Antenna In 
This field is used to indicate losses or gains between the ANT IN port 
and the device-under-test. 
Enter a positive value to indicate a gain (such as an amplifier). The 
spectrum analyzer’s marker level (
Lvl
) measurement is automatically 
reduced by that amount. The spectrum analyzer’s 
Ref Level
 is 
automatically decreased by the same amount, so the trace position does 
not appear to change. 
Enter a negative value to indicate a loss (such as cable loss). The 
spectrum analyzer marker’s level (
Lvl
) measurement is automatically 
increased by that amount. The spectrum analyzer’s 
Ref Level
 is 
automatically increased by the same amount, so the trace position does 
not appear to change.
This field is only used when the 
RF Level Offset
 field is set to 
On
Se
Screen(s) Where Field is Present
INSTRUMENT CONFIGURE