Chapter 6 
Configuration - Screens and Control Fields
Flow Control
Flow Control
Flow Control
The flow control field is used when sending or receiving serial data. 
None
 disables the Xon/Xoff function. 
Xon/Xoff
 lets the Test Set “talk/listen” to the transceiver to alter 
the rate of the data being sent. 
Hardware
 uses the serial port’s RTS and CTS lines to transfer data 
at high baud rates.
NOTE
Do not turn flow control on (set 
Flow Cntl 
to 
Hardware
 
in the I/O CONFIGURE screen) for the serial port until 
you have a device attached to the port that can respond to 
the flow control communications.
Example: You have a printer attached to SERIAL PORT 9. 
Before you run any IBASIC software (for example, the 
RFTOOLS program), you must make sure that the printer 
is properly connected and that its power is turned on.
Operating Considerations
This setting is retained when the instrument is turned off. 
Baud rates of 38400 and above must use 
Hardware
 flow control.
Screen(s) Where Field is Present
I/O CONFIGURE
Main Menu