Connectors
RF IN/OUT
SCOPE MONITOR OUTPUT
RF IN/OUT 
This Type-N connection is used to output signals from the RF generator 
and to input RF signals. 
CAUTION
Overpower Damage — Refer to the Test Set’s connector 
panel for maximum input power level. Exceeding this level 
can cause permanent instrument damage. 
If the RF power at the RF IN/OUT port exceeds allowable 
limits, a loud warning signal sounds and a message 
appears at the top of the screen. If this occurs, disconnect 
the RF power, reset using 
Meas Reset
 (press and release the 
Shift
 key, then the 
Hold
 key), and allow the Test Set to cool 
off for approximately 2 minutes before making any other 
measurements on this port.
Operating Considerations
This port must be used when measuring TX (RF) power. 
Signals 
60 mW can be input to the ANT IN connector for all RF 
measurements except TX Power. 
This port can be selected on the RF ANALYZER, or SPEC ANL screens. 
SCOPE MONITOR OUTPUT 
This connector provides an external output from the AF analyzer. 
Operating Considerations
The 
Scope To
 field in the AF ANALYZER screen determines the 
source of this signal. For more information see the 
.
The level is not affected by the front-panel VOLUME knob. 