Chapter 2 
Operating Overview
Averaging Measurements
Averaging Measurements 
The 
Avg
 (average) function allows you to display the average value of a 
number of measurements. You enter the number of measurement 
samples used to calculate and display the measurement average. This 
dampens the effects of rapidly changing measurements, providing a 
more usable measurement display. 
To use measurement averaging 
Step 1. Position the cursor in front of the measurement’s 
unit-of-measure. 
Step 2. Press and release the 
Shift
 key, then the 
Incr 
×
10
 key to access 
the 
Avg
 function. 
Step 3. Enter the desired number of measurement samples to be used 
for calculating the average, then press the knob or the 
Enter
 
key.
or
Press 
On/Off
 to use the currently-displayed number of samples.
Step 4. To turn averaging off, position the cursor in front of the 
unit-of-measure and press and release the 
Shift
 key, then the 
Incr 
×
10
 key to access the 
Avg
 function, then press the 
On/Off
 key. 
How the Test Set Averages Measurements
When the averaging function is first enabled, a numeric average is 
calculated and displayed each time a measurement is made. This 
continues until the specified number of samples is reached. From that 
point on, the averaging function performs an exponential filtering 
operation that mimics an RC filter. 
Because of the exponential response, any large measurement changes 
result in a displayed value that ramps up or down to the actual 
measured value. 
A measurement reset clears the measurement history for all 
measurements and starts the averaging process over. (Press and 
release the 
Shift
 key then the 
Hold
 key (
Meas Reset
) to reset the 
measurement.) 
Main Menu