Справочник Пользователя для Pioneer Q150

Скачать
Страница из 17
 
14
4.  The oxygen concentration should be checked every 90 days for units without optional oxygen 
sensors. The oxygen concentration should be checked every 6 months for units with optional 
oxygen sensors.   
 
10.2 CLEANING THE CABINET FILTERS 
NOTE: All concentrator models have one cabinet filter located at rear side of the cabinet. 
(FIG. 6) 
 
 
FIGURE 6 
  CABINET FILTER LOCATION 
 
1.  Remove the cabinet filter and 
CLEAN AT LEAST ONCE A WEEK depending on environmental 
conditions. 
(FIG.7) 
2.  Clean the cabinet filter with a vacuum cleaner or wash in warm soapy water and rinse thoroughly. 
(FIG. 7) 
3.  Dry the filters thoroughly before reinstallation. 
 
 
FIGURE 7 
  REMOVE THE CABINET FILTERS 
 
 
10.3 CABINET 
Clean the cabinet with a mild household cleaner and non-abrasive cloth or sponge.